HTML has links - PDF has Authentication
246-240-611  <<  246-240-614 >>   246-240-617

PDFWAC 246-240-614

Records of teletherapy, remote afterloader, and gamma stereotactic radiosurgery full calibrations.

(1) A licensee shall maintain a record of the teletherapy unit, remote afterloader unit, and gamma stereotactic radiosurgery unit full calibrations required by WAC 246-240-369, 246-240-372, and 246-240-375 for three years.
(2) The record must include:
(a) The date of the calibration;
(b) The manufacturer's name, model number, and serial number of the teletherapy, remote afterloader, and gamma stereotactic radiosurgery units, the sources, and the instruments used to calibrate the units;
(c) The results and an assessment of the full calibrations;
(d) The results of the autoradiograph required for low dose-rate remote afterloader units; and
(e) The signature of the authorized medical physicist who performed the full calibration.
[Statutory Authority: RCW 70A.388.040 and 70A.388.110. WSR 22-19-084, § 246-240-614, filed 9/20/22, effective 10/21/22. Statutory Authority: RCW 70.98.050. WSR 06-05-019, § 246-240-614, filed 2/6/06, effective 3/9/06.]