HTML has links - PDF has Authentication
246-226-040  <<  246-226-050 >>   246-226-060

PDFWAC 246-226-050

Dose limits.

The CTDIvol for the following CT procedure on phantoms may not exceed the dose limits established in Table 1.
Table 1: Dose limits
CT Procedure
Phantom Size
Dose Limit:
CTDIvol(mGy)
Adult head
16 cm
80
Adult abdomen
32 cm
30
Pediatric head (one year old)
16 cm
40
Pediatric abdomen
(40 pounds)
16 cm
20
[Statutory Authority: RCW 70.98.050 and 70.98.080. WSR 16-23-030, ยง 246-226-050, filed 11/8/16, effective 1/1/17.]
Site Contents
Selected content listed in alphabetical order under each group